General Alarms and Troubleshooting
12-48
COBE
®
Spectra™ Apheresis System •
Essentials Guide
PLASMA VALVE NOT OPERATING CORRECTLY!
Alarm LED is red. Pumps stop.
To retry the plasma valve test, press
CONTINUE
.
This alarm occurs only during TPE and RBCX procedures when the
plasma pump fails to generate expected pressures at the return pressure
sensor for various plasma valve positions.
PLASMA VALVE NOT OPERATING CORRECTLY!
Check for proper loading.
CONTINUE
Probable Cause
Action
Tubing incorrectly positioned in
plasma valve
Ensure the tubing is placed correctly in the plasma valve. Press
CONTINUE
.
Replacement fluid or collection bag
lines clamped
Ensure the replacement or collection bag lines are unclamped. Press
CONTINUE
.
Leak at the return pressure sensor
pod seal
Check the return pressure sensor pod seal for wear or damage. Replace if necessary. For instructions, see “Changing pressure pod
seals” on page 10-4.
Note:
For instructions on repositioning the diaphragm and reloading the return pressure sensor, see “Repositioning the
Return Pressure Sensor Diaphragm” on page 11-10.
Return Flow Controller incorrectly set
Ensure the Return Flow Controller is completely in the prime position; or during Run mode, it is in the proper setting.
Plasma valve in the wrong position
Note the alarm message. Notify a qualified service technician of the alarm message.
If alarm occurs during the Run mode of a procedure, use a flat-blade screwdriver to manually place the valve in the proper position.
Pump malfunction
Note the alarm message. Discontinue the procedure. Notify a qualified service technician of the alarm message.
Summary of Contents for Cobe Spectra
Page 1: ......
Page 7: ...COBE Spectra Apheresis System Essentials Guide 1 1 1 Preface ...
Page 27: ...COBE Spectra Apheresis System Essentials Guide 2 1 2 Introduction ...
Page 75: ...COBE Spectra Apheresis System Essentials Guide 5 1 5 Software Options and Configuration ...
Page 92: ...Software Options and Configuration 5 18 COBE Spectra Apheresis System Essentials Guide ...
Page 93: ...COBE Spectra Apheresis System Essentials Guide 6 1 6 Automatic and Manual Modes ...
Page 98: ...Automatic and Manual Modes 6 6 COBE Spectra Apheresis System Essentials Guide ...
Page 99: ...COBE Spectra Apheresis System Essentials Guide 7 1 7 Alarm Tests ...
Page 103: ...COBE Spectra Apheresis System Essentials Guide 8 1 8 Anticoagulation ...
Page 109: ...COBE Spectra Apheresis System Essentials Guide 9 1 9 Fluid Volumes ...
Page 115: ...COBE Spectra Apheresis System Essentials Guide 10 1 10 General Maintenance ...
Page 120: ...General Maintenance 10 6 COBE Spectra Apheresis System Essentials Guide ...
Page 121: ...COBE Spectra Apheresis System Essentials Guide 11 1 11 Helpful Hints and Recovery Procedures ...
Page 133: ...COBE Spectra Apheresis System Essentials Guide 12 1 12 General Alarms and Troubleshooting ...
Page 197: ...COBE Spectra Apheresis System Essentials Guide 13 1 13 Specifications ...
Page 208: ...Specifications 13 12 COBE Spectra Apheresis System Essentials Guide ...
Page 209: ...COBE Spectra Apheresis System Essentials Guide 14 1 14 Data Input Limits ...
Page 212: ...Data Input Limits 14 4 COBE Spectra Apheresis System Essentials Guide ...
Page 213: ...COBE Spectra Apheresis System Essentials Guide 15 1 15 The Seal Safe System ...
Page 224: ...The Seal Safe System 15 12 COBE Spectra Apheresis System Essentials Guide ...
Page 225: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Page 226: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Page 227: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Page 228: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Page 229: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Page 230: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Page 231: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Page 232: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Page 233: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Page 234: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...