General Alarms and Troubleshooting
COBE
®
Spectra™ Apheresis System •
Essentials Guide
12-25
Check the CCM for proper loading
Alarm LED is red.
Press
CONTINUE
to remove the alarm message.
.
COLLECT PUMP ERROR!
Alarm LED is red. Pumps stop.
To retry the collect pump test, press
CONTINUE
.
This alarm occurs only during Prime mode when the collect pump fails
to develop sufficient pressure at the return pressure sensor.
Check the CCM for proper loading
Press CONTINUE
Probable Cause
Action
Cuvette not properly installed in the CCM
Ensure the cuvette is properly installed in the CCM. Press
CONTINUE
.
Incorrect disposable tubing set selected during data entry
Press
CHANGE MODE
, press
1
to select “Load Set,” choose the correct disposable tubing set, and reprime.
COLLECT PUMP ERROR! Check for proper
loading or leaks.
CONTINUE
Probable Cause
Action
Tubing incorrectly positioned in the collect pump or return
pressure sensor
Ensure the collect pump, valves, and return pressure sensor are loaded correctly.
Note:
For instructions on how to reposition the diaphragm and reload the return pressure sensor,
see “Repositioning the Return Pressure Sensor Diaphragm” on page 11-10.
Press
CONTINUE
.
Plasma, return, waste, RBC, or collect valve incorrectly
loaded
Check for proper valve loading.
Incorrect disposable tubing set selected during data entry
Press
CHANGE MODE
, press
1
to select “Load Set,” choose the correct disposable tubing set, then reprime.
Summary of Contents for Cobe Spectra
Page 1: ......
Page 7: ...COBE Spectra Apheresis System Essentials Guide 1 1 1 Preface ...
Page 27: ...COBE Spectra Apheresis System Essentials Guide 2 1 2 Introduction ...
Page 75: ...COBE Spectra Apheresis System Essentials Guide 5 1 5 Software Options and Configuration ...
Page 92: ...Software Options and Configuration 5 18 COBE Spectra Apheresis System Essentials Guide ...
Page 93: ...COBE Spectra Apheresis System Essentials Guide 6 1 6 Automatic and Manual Modes ...
Page 98: ...Automatic and Manual Modes 6 6 COBE Spectra Apheresis System Essentials Guide ...
Page 99: ...COBE Spectra Apheresis System Essentials Guide 7 1 7 Alarm Tests ...
Page 103: ...COBE Spectra Apheresis System Essentials Guide 8 1 8 Anticoagulation ...
Page 109: ...COBE Spectra Apheresis System Essentials Guide 9 1 9 Fluid Volumes ...
Page 115: ...COBE Spectra Apheresis System Essentials Guide 10 1 10 General Maintenance ...
Page 120: ...General Maintenance 10 6 COBE Spectra Apheresis System Essentials Guide ...
Page 121: ...COBE Spectra Apheresis System Essentials Guide 11 1 11 Helpful Hints and Recovery Procedures ...
Page 133: ...COBE Spectra Apheresis System Essentials Guide 12 1 12 General Alarms and Troubleshooting ...
Page 197: ...COBE Spectra Apheresis System Essentials Guide 13 1 13 Specifications ...
Page 208: ...Specifications 13 12 COBE Spectra Apheresis System Essentials Guide ...
Page 209: ...COBE Spectra Apheresis System Essentials Guide 14 1 14 Data Input Limits ...
Page 212: ...Data Input Limits 14 4 COBE Spectra Apheresis System Essentials Guide ...
Page 213: ...COBE Spectra Apheresis System Essentials Guide 15 1 15 The Seal Safe System ...
Page 224: ...The Seal Safe System 15 12 COBE Spectra Apheresis System Essentials Guide ...
Page 225: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Page 226: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Page 227: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Page 228: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Page 229: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Page 230: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Page 231: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Page 232: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Page 233: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Page 234: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...