General Alarms and Troubleshooting
12-22
COBE
®
Spectra™ Apheresis System •
Essentials Guide
CENTRIFUGE PRESSURE ERROR!
Alarm LED is red. Pumps stop.
To retry the centrifuge pressure sensor test, press
CONTINUE
.
This alarm occurs only during Prime mode if the centrifuge pressure
sensor does not detect a pressure increase with the RBC and waste valves
closed.
.
CENTRIFUGE PRESSURE ERROR!
Check for proper loading.
CONTINUE
Probable Cause
Action
Tubing incorrectly positioned in the centrifuge pressure
sensor
Ensure the tubing is correctly and completely positioned in the centrifuge pressure sensor. Press
CONTINUE
.
If the alarm recurs, reprime the disposable tubing set by pressing
CHANGE MODE
, and then pressing
2
to
select “Prime.”
RBC or waste valve not loaded
Ensure the tubing is placed in the RBC and waste valves. Reprime the disposable tubing set.
Leak in the disposable tubing set
Check the disposable tubing set for leaks. If the set does not leak, press
CONTINUE
. If the set leaks, replace it
with a new set and prime the set.
If the alarm recurs and no machine malfunction is indicated, replace the disposable tubing set.
Sensor, valve, or pump malfunction
Note the alarm message. Discontinue the procedure. Notify a qualified service technician of the alarm
message.
Summary of Contents for Cobe Spectra
Page 1: ......
Page 7: ...COBE Spectra Apheresis System Essentials Guide 1 1 1 Preface ...
Page 27: ...COBE Spectra Apheresis System Essentials Guide 2 1 2 Introduction ...
Page 75: ...COBE Spectra Apheresis System Essentials Guide 5 1 5 Software Options and Configuration ...
Page 92: ...Software Options and Configuration 5 18 COBE Spectra Apheresis System Essentials Guide ...
Page 93: ...COBE Spectra Apheresis System Essentials Guide 6 1 6 Automatic and Manual Modes ...
Page 98: ...Automatic and Manual Modes 6 6 COBE Spectra Apheresis System Essentials Guide ...
Page 99: ...COBE Spectra Apheresis System Essentials Guide 7 1 7 Alarm Tests ...
Page 103: ...COBE Spectra Apheresis System Essentials Guide 8 1 8 Anticoagulation ...
Page 109: ...COBE Spectra Apheresis System Essentials Guide 9 1 9 Fluid Volumes ...
Page 115: ...COBE Spectra Apheresis System Essentials Guide 10 1 10 General Maintenance ...
Page 120: ...General Maintenance 10 6 COBE Spectra Apheresis System Essentials Guide ...
Page 121: ...COBE Spectra Apheresis System Essentials Guide 11 1 11 Helpful Hints and Recovery Procedures ...
Page 133: ...COBE Spectra Apheresis System Essentials Guide 12 1 12 General Alarms and Troubleshooting ...
Page 197: ...COBE Spectra Apheresis System Essentials Guide 13 1 13 Specifications ...
Page 208: ...Specifications 13 12 COBE Spectra Apheresis System Essentials Guide ...
Page 209: ...COBE Spectra Apheresis System Essentials Guide 14 1 14 Data Input Limits ...
Page 212: ...Data Input Limits 14 4 COBE Spectra Apheresis System Essentials Guide ...
Page 213: ...COBE Spectra Apheresis System Essentials Guide 15 1 15 The Seal Safe System ...
Page 224: ...The Seal Safe System 15 12 COBE Spectra Apheresis System Essentials Guide ...
Page 225: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Page 226: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Page 227: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Page 228: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Page 229: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Page 230: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Page 231: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Page 232: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Page 233: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Page 234: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...