Introduction
COBE
®
Spectra™ Apheresis System •
Essentials Guide
2-13
3
Press
MENU ON/OFF
to exit the menu system when the Menu
LED illuminates .
4
Press
CLEAR
to back out of the menu system one menu at a time.
For more information, see “The MENU ON/OFF Key” on page 5-8.
Pause LED
(green LED above the PAUSE/CONTINUE key):
Illuminates to indicate that the Spectra system is paused and the pumps
are stopped.
PAUSE/CONTINUE key:
1
Press
PAUSE/CONTINUE
to pause the Spectra system and stop
the pumps. The Pause LED illuminates. If the pumps stop for more
than 60 seconds, the centrifuge speed slows to 1800 rpm to reduce
the temperature in the channel’s multi-lumen tubing.
2
Press
PAUSE/CONTINUE
when the Pause LED flashes to restart
the pumps and turn off the Pause LED. The centrifuge speed
increases automatically if it previously slowed to 1800 rpm.
Certain alarm conditions cause the Spectra system to pause. The
system prompts you to clear the alarm (the Pause LED is steady, not
flashing) before pressing
PAUSE/CONTINUE
.
Display
The control panel includes a 2-line, 40-character display. Displayed
information falls into six general categories:
•
State screens
•
Alarm screens
•
Operator information screens
•
VALVE key displays
•
CHANGE MODE key displays
•
MENU ON/OFF key displays
State screens:
Indicate the current state of the Spectra system. (A state is
one of many individual steps that the Spectra system performs to
complete each mode of operation.) The step determines the pump flow
rates, centrifuge speed, valve positions, enabled alarms, and display
screens. The screen explains what action the system performs, or
prompts you to take a specific action or make a selection from a menu.
The current mode of operation or a procedure abbreviation appears in
the lower right-hand corner of the screen. A list of these abbreviations
follows this section.
Alarm screens:
Identify the source of an alarm and prompt you with the
action to take to clear the alarm condition. When there is more than one
alarm, the highest priority alarm screen appears first and has an asterisk
(*) in the lower right-hand corner. This indicates there are more alarms
than the one that currently appears. Press
CLEAR
to temporarily view
the lower-priority alarm(s). The highest-priority alarm reappears after
30 seconds or when you press
ENTER
. For additional information on
Spectra alarms and the actions to take in response to those alarms, see
Chapter 12, “General Alarms and Troubleshooting,” or the alarms and
troubleshooting sections in the procedural guides.
Summary of Contents for Cobe Spectra
Page 1: ......
Page 7: ...COBE Spectra Apheresis System Essentials Guide 1 1 1 Preface ...
Page 27: ...COBE Spectra Apheresis System Essentials Guide 2 1 2 Introduction ...
Page 75: ...COBE Spectra Apheresis System Essentials Guide 5 1 5 Software Options and Configuration ...
Page 92: ...Software Options and Configuration 5 18 COBE Spectra Apheresis System Essentials Guide ...
Page 93: ...COBE Spectra Apheresis System Essentials Guide 6 1 6 Automatic and Manual Modes ...
Page 98: ...Automatic and Manual Modes 6 6 COBE Spectra Apheresis System Essentials Guide ...
Page 99: ...COBE Spectra Apheresis System Essentials Guide 7 1 7 Alarm Tests ...
Page 103: ...COBE Spectra Apheresis System Essentials Guide 8 1 8 Anticoagulation ...
Page 109: ...COBE Spectra Apheresis System Essentials Guide 9 1 9 Fluid Volumes ...
Page 115: ...COBE Spectra Apheresis System Essentials Guide 10 1 10 General Maintenance ...
Page 120: ...General Maintenance 10 6 COBE Spectra Apheresis System Essentials Guide ...
Page 121: ...COBE Spectra Apheresis System Essentials Guide 11 1 11 Helpful Hints and Recovery Procedures ...
Page 133: ...COBE Spectra Apheresis System Essentials Guide 12 1 12 General Alarms and Troubleshooting ...
Page 197: ...COBE Spectra Apheresis System Essentials Guide 13 1 13 Specifications ...
Page 208: ...Specifications 13 12 COBE Spectra Apheresis System Essentials Guide ...
Page 209: ...COBE Spectra Apheresis System Essentials Guide 14 1 14 Data Input Limits ...
Page 212: ...Data Input Limits 14 4 COBE Spectra Apheresis System Essentials Guide ...
Page 213: ...COBE Spectra Apheresis System Essentials Guide 15 1 15 The Seal Safe System ...
Page 224: ...The Seal Safe System 15 12 COBE Spectra Apheresis System Essentials Guide ...
Page 225: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Page 226: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Page 227: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Page 228: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Page 229: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Page 230: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Page 231: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Page 232: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Page 233: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Page 234: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...