Preface
1-8
COBE
®
Spectra™ Apheresis System •
Essentials Guide
Warning:
The Spectra system alarm system is inactive during
manual Rinseback. If you must perform a manual Rinseback,
watch for air in the return line when returning blood to the donor or
patient. If you see air, immediately discontinue the Rinseback.
Warning:
Inspect blood components for bacterial growth before
transfusion according to your local transfusion standards (for
example, the American Association of Blood Banks’ Standards for
Blood Banks and Transfusion Services).
Warning:
If you choose to perform the alarm tests, DO NOT
connect the donor or patient until the alarm tests are complete.
Warning:
DO NOT use the AutoPBSC disposable tubing set if the
access needle is disconnected from the disposable tubing set
when you open the package. The disposable tubing set may no
longer be considered sterile or functionally closed.
Warning:
DO NOT use the ELP or LRS disposable tubing set if
the access needle is disconnected from the disposable tubing set
when you open the package. The disposable tubing set may no
longer be considered sterile or functionally closed.
Warning:
DO NOT use the ELP, LRS, AutoPBSC, or Functionally
Closed WBC disposable tubing set if the access needle is
disconnected from the disposable tubing set when you open the
package.The disposable tubing set may no longer be considered
sterile or functionally closed.
Warning:
The Functionally Closed WBC disposable tubing set is
no longer functionally closed if
•
The first attempt to successfully insert the access needle fails
(unless you attach a new needle using an approved
connection device which maintains the sterility of the
disposable tubing set).
•
The access needle is disconnected during the procedure.
•
You use the injection site on the plasma line to obtain blood
samples or to infuse medication or parenteral solutions.
•
You prime the disposable tubing set with any anticoagulant
other than ACD-A.
•
You disconnect any product bag before you seal it.
•
The integrity of the disposable tubing set is compromised for
any reason.
Warning:
The AutoPBSC disposable tubing set is no longer
functionally closed if
•
The first attempt to successfully insert the access needle fails
(unless you attach a new needle using an approved
connection device which maintains the sterility of the
disposable tubing set).
•
The access needle is disconnected during the procedure.
•
You disconnect any product bag before you seal it.
•
The integrity of the disposable tubing set is compromised for
any reason.
Summary of Contents for Cobe Spectra
Page 1: ......
Page 7: ...COBE Spectra Apheresis System Essentials Guide 1 1 1 Preface ...
Page 27: ...COBE Spectra Apheresis System Essentials Guide 2 1 2 Introduction ...
Page 75: ...COBE Spectra Apheresis System Essentials Guide 5 1 5 Software Options and Configuration ...
Page 92: ...Software Options and Configuration 5 18 COBE Spectra Apheresis System Essentials Guide ...
Page 93: ...COBE Spectra Apheresis System Essentials Guide 6 1 6 Automatic and Manual Modes ...
Page 98: ...Automatic and Manual Modes 6 6 COBE Spectra Apheresis System Essentials Guide ...
Page 99: ...COBE Spectra Apheresis System Essentials Guide 7 1 7 Alarm Tests ...
Page 103: ...COBE Spectra Apheresis System Essentials Guide 8 1 8 Anticoagulation ...
Page 109: ...COBE Spectra Apheresis System Essentials Guide 9 1 9 Fluid Volumes ...
Page 115: ...COBE Spectra Apheresis System Essentials Guide 10 1 10 General Maintenance ...
Page 120: ...General Maintenance 10 6 COBE Spectra Apheresis System Essentials Guide ...
Page 121: ...COBE Spectra Apheresis System Essentials Guide 11 1 11 Helpful Hints and Recovery Procedures ...
Page 133: ...COBE Spectra Apheresis System Essentials Guide 12 1 12 General Alarms and Troubleshooting ...
Page 197: ...COBE Spectra Apheresis System Essentials Guide 13 1 13 Specifications ...
Page 208: ...Specifications 13 12 COBE Spectra Apheresis System Essentials Guide ...
Page 209: ...COBE Spectra Apheresis System Essentials Guide 14 1 14 Data Input Limits ...
Page 212: ...Data Input Limits 14 4 COBE Spectra Apheresis System Essentials Guide ...
Page 213: ...COBE Spectra Apheresis System Essentials Guide 15 1 15 The Seal Safe System ...
Page 224: ...The Seal Safe System 15 12 COBE Spectra Apheresis System Essentials Guide ...
Page 225: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Page 226: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Page 227: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Page 228: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Page 229: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Page 230: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Page 231: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Page 232: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Page 233: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Page 234: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...