5322A
Calibration Manual
2
Safety Information
General Safety Information is in the printed Safety Information document that ships with the
Product and is available at
. More specific safety information is listed in this
manual where applicable.
A
Warning
identifies hazardous conditions and procedures that are dangerous to the user. A
Caution
identifies conditions and procedures that can cause damage to the Product or the
equipment under test.
Specifications
Safety Specifications are in the Safety Specifications section of the
Safety Information
document.
Complete Product specifications are at
www.flukecal.com
. See the
5322A Product
Specifications
.
Theory of Operation
The Product is a Multifunction Electrical Tester Calibrator that covers a wide range of
calibrations of devices under test (DUT). The Product produces calibrated outputs of these
electrical quantities:
• Ground bond resistance source in fixed values from 25 m
Ω
to 1.8 k
Ω
• Low-signal level, low-resistance source from 100 m
Ω
to 10 k
Ω
with 3 1/2-digit resolution
• High-resistance source in range from 10 k
Ω
to 10 G
Ω
with 4 1/2-digit resolution plus a
single 100 G
Ω
value
• AC/DC voltage calibrator
• Loop/Line impedance calibrator that uses a ground bond resistance decade and residual
impedance compensator (optional)
• RCD calibrator in a range from 3 mA to 3 A with trip time from 10 ms to 5 s
• Leakage current source, both passive and active (optional)
• AC/DC multimeter to 1100 V (10kV or 40 kV with an external probe) and 30 A