5322A
Calibration Manual
44
Table 9. AC Voltage Calibrator Adjustment Points
8. Highlight the
0.03 V 55Hz low (30 mV)
adjustment point and push the
Select
softkey. The
Product shows the value entered in the last adjustment.
9. Use the cursor keys to set the value to the value shown on the standard multimeter:
30.00 mV.
10. Push the
Write
softkey to save the new value.
11. Do the same procedure all adjustment points.
12. Push the
Exit
softkey to go to a higher level in menu.
This completes this part of the adjustment.
Nominal Voltage
Range
Adjustment Point
Frequency
Required Accuracy of
Adjustment
0.3 V ac
0.03 V ac low
55 Hz
0.03 %
0.3 V ac
0.3 V ac high
55 Hz
0.03 %
3V ac
0.3 V ac low
55 Hz
0.03 %
3V ac
3V ac high
55 Hz
0.03 %
30 V ac
3V ac low
55 Hz
0.03 %
30 V ac
30V ac high
55 Hz
0.03 %
150 V ac
30 V ac low
55 Hz
0.03 %
150 V ac
150 V ac high
55 Hz
0.03 %
300 V ac
150 V ac low
55 Hz
0.03 %
300 V ac
300 V ac high
55 Hz
0.03 %
600 V ac
300 V ac low
55 Hz
0.03 %
600 V ac
600 V ac high
55 Hz
0.03 %