5322A
Calibration Manual
42
DC Voltage Calibrator Adjustment (Optional)
Calibration consists of adjustments of these ranges: 0.3 V, 3 V, 30 V, 150 V, and 600 V. Each of
these has two adjustment points.
To adjust the dc voltage parameter, see
1. Connect a standard multimeter to the Product output terminals.
Figure 31. DC Voltage Adjustment Connections
2. On the Product, from the Setup menu, set the grounding mode to
GND ON
.
3. Set the standard multimeter to the most accurate dc voltage mode.
4. Push the
Setup
softkey on the Product.
5. Highlight
Calibration
and push the
Select
softkey.
6. From the calibration menu, highlight
DCV Calibrator
and push the
Select
softkey. A list of
internal ranges shows.
Fluke 5322A
Fluke 8588A