5322A
Calibration Manual
16
For example, in
, the Ground Bond Resistance screen shows the resistance
segment to adjust with this syntax:
• GBR
ground bond resistance decade
• R02
ordinal number of the resistance segment
• 100 m
Ω
nominal value of calibrated resistance segment
• 102.320 m
Ω
adjustment value of R02 segment
Figure 14. Ground Bond Resistance Decade
2. The Product measures points where the external standard signal is applied to the Product
input terminals. Use the cursor keys or the rotary knob to set the value on the display to the
value shown on the external standard source. Although a preferred adjustment point is
prescribed in the display, any suitable value within the range of the meter is applicable.
These functions use this system:
• High Resistance Meters
• Low Resistance Meters
• Ground Bond Resistance Meters
• RCD Trip Current
• Leakage Current