Multifunction Electrical Tester Calibrator
Adjustment Procedures
35
To set the required constants to adjust the test voltage auxiliary voltmeter:
1. Connect the 5522A to the OUTPUT HI
Ω
HI and LO terminals of the Product, see
Figure 26. Low Resistance Source Meters Adjustment Connections 2
2. On the Product, highlight
Low Resistance Meters
from calibration menu and push the
Select
softkey.
3. Select
LVR V01 Voltage 1Vdc
and push the
Select
softkey.
4. Set the output voltage to 1 V dc on the 5522A and switch the output terminals ON.
5. On the Product, use the rotary knob or cursor keys to set the value to 45.0 V.
6. Push the
Write
softkey to confirm the new adjustment value.
7. Repeat step 6 for the next three levels: V03 5 V dc, V04 15 V dc, and V05 50 V dc. Use the
dc output voltage from the 5522A.
This completes this part of the adjustment.
1
4
7
/
+
STBY
5522A CALIBRATOR
Fluke 5322A Fluke 5522A