Multifunction Electrical Tester Calibrator
Adjustment Procedures
31
Table 6. High Resistance Source Adjustment for 1.5 kV Adjustment Points
Nominal Value
Position
Required
Accuracy of
Adjustment
Lower Limit
Upper Limit
10 k
Ω
HVR00
0.02 %
-5 %
+5 %
18 k
Ω
HVR01
0.02 %
-5 %
+5 %
33 k
Ω
HVR02
0.02 %
-5 %
+5 %
44 k
Ω
HVR03
0.02 %
-5 %
+5 %
100 k
Ω
HVR04
0.02 %
-5 %
+5 %
180 k
Ω
HVR05
0.02 %
-5 %
+5 %
340 k
Ω
HVR06
0.02 %
-5 %
+5 %
480 k
Ω
HVR07
0.02 %
-5 %
+5 %
930 k
Ω
HVR08
0.05 %
-5 %
+5 %
1.9 M
Ω
HVR09
0.05 %
-5 %
+5 %
3.6 M
Ω
HVR10
0.05 %
-5 %
+5 %
3.5 M
Ω
HVR11
0.05 %
-5 %
+5 %
10.7 M
Ω
HVR12
0.1 %
-10 %
+10 %
20.8 M
Ω
HVR13
0.1 %
-10 %
+10 %
40.6 M
Ω
HVR14
0.1 %
-10 %
+10 %
40.6 M
Ω
HVR15
0.1 %
-10 %
+10 %
103 M
Ω
HVR16
0.2 %
-10 %
+10 %
206 M
Ω
HVR17
0.2 %
-10 %
+10 %
400 M
Ω
HVR18
0.2 %
-10 %
+10 %
390 M
Ω
HVR19
0.2 %
-10 %
+10 %
1.0 G
Ω
HVR20
0.5 %
-10 %
+10 %
1.8 G
Ω
HVR21
0.5 %
-10 %
+10 %
2.7 G
Ω
HVR22
0.5 %
-10 %
+10 %
5.6 G
Ω
HVR23
0.5 %
-10 %
+10 %
100 G
Ω
HVR24
1.0 %
-10 %
+10 %