Multifunction Electrical Tester Calibrator
Verification
87
Nominal Output
Voltage
Frequency
(Hz)
Lower Limit
(A)
Upper Limit
(A)
2.5 mA
400
2.4935 mA
2.5065 mA
10 mA
50
9.965 mA
10.035 mA
20 mA
50
19.945 mA
20.055 mA
25 mA
50
24.935 mA
25.065 mA
25 mA
60
24.935 mA
25.065 mA
25 mA
400
24.935 mA
25.065 mA
100 mA
50
99.65 mA
100.35 mA
200 mA
50
199.45 mA
200.55 mA
250 mA
50
249.35 mA
250.65 mA
250 mA
60
249.35 mA
250.65 mA
250 mA
400
249.35 mA
250.65 mA
Table 33. DC/AC Voltage Multimeter Test 5 kV Option
Nominal Output
Voltage
Frequency
(Hz)
Lower Limit
(V)
Upper Limit
(V)
1000 V
-
991.5
1008.5
2000 V
-
1988.5
2011.5
3000 V
-
2985.5
3014.5
4000 V
-
3982.5
4017.5
5000 V
-
4979.5
5020.5
1000 V
50
991.5
1008.5
2000 V
50
1988.5
2011.5
3000 V
50
2985.5
3014.5
4000 V
50
3982.5
4017.5
5000 V
50
4979.5
5020.5
1000 V
60
991.5
1008.5
2000 V
60
1988.5
2011.5
3000 V
60
2985.5
3014.5
4000 V
60
3982.5
4017.5
5000 V
60
4979.5
5020.5
1000 V
0
991.5
1008.5
Table 32. DC/AC HIPOT Leakage Current Test (cont.)