5322A
Calibration Manual
70
Voltage Calibrator
Follow these steps to verify the voltage calibrator functions.
1. Voltage calibrator test:
a. On the Product, select the dc voltage calibrator function.
b. Connect the 5522A to the appropriate output terminals of the Product and select the
dc voltage function.
. Do not exceed the specified limits.
d. Select the ac voltage calibrator function and set the output frequency 55 Hz.
e. Connect the standard multimeter to the appropriate output terminals of the Product
and select the ac voltage function.
f. Do the ac voltage test, see
. Do not exceed the specified limits.
2. Frequency calibrator test:
a. Select ac voltage calibrator function on the Product.
b. Set the value to 10 V at 400 Hz.
c. Connect the standard counter (Keysight 53131A) to the HI – LO output terminals of the
Product and select
Frequency
measurement
.
d. Push
O
on the Product.
e. The 53131A starts to measure the voltage calibrator frequency. Do not exceed the
limits of 399.92 Hz to 400.08 Hz.