Multifunction Electrical Tester Calibrator
Calibration
17
For example, in
, the display indicates the High Resistance Source Voltmeter
segment to be adjusted in this syntax:
• HVR
High voltage decade
• V01
Voltmeter senses voltage on the decade, range 01
• Measured value Life reading of the High Resistance Voltmeter
• 1368.57
Adjustable calibration value. Set such value to get reading which
equals the set output voltage on the external standard.
Figure 15. High Resistance Voltage Screen
3. Points where the Product measures the external standard signal on the input terminals. Do
not manually adjust this signal.
These functions use this system:
• DC Multimeter
• AC Multimeter