5322A
Calibration Manual
38
Table 7. Adjustment - High Resistance Meters 5 kV Adjustment
Ammeter Adjustment HVR I01
One constant must be set to 20 mA dc for the current range. To adjust the Product:
1. Connect the 5522A to the OUTPUT HI
Ω
HI and LO terminals of the Product, see
2. Select
HVR I01 Current 20mAdc (Short)
from the calibration menu and push
Select
.
3. On the 5522A, set the output current to 20 mA dc and switch the output terminals ON.
4. Use the rotary knob or cursor keys to set the Product to 20.0 mA.
5. Push the
Write
softkey to confirm the new adjustment value.
Adjustment Point
Output Voltage
Resistance Condition
[1]
Calibration Constant
HVR V01
0.0 V dc
1 M
Ω
Offset
HVR V01
1100 V dc
1 M
Ω
Slope
HVR V02
0.0 V dc
10 M
Ω
Offset
HVR V02
2500 V dc
10 M
Ω
Slope
HVR V03
0.0 V dc
1.4 G
Ω
Offset
HVR V03
6000 V dc
1.4 G
Ω
Slope
HVR V04
0.0 V dc
2.5 G
Ω
Offset
HVR V04
6000 V dc
2.5 G
Ω
Slope
HVR V05
0.0 V dc
5 G
Ω
Offset
HVR V05
6000 V dc
5 G
Ω
Slope
HVR V06
0.0 V dc
10 G
Ω
Offset
HVR V06
6000 V dc
10 G
Ω
Slope
HVR V07
0.0 V dc
22 G
Ω
Offset
HVR V07
6000 V dc
22 G
Ω
Slope
HVR V08
0.0 V dc
50 G
Ω
Offset
HVR V08
6000 V dc
50 G
Ω
Slope
HVR V09
0.0 V dc
100 G
Ω
Offset
HVR V09
6000 V dc
100 G
Ω
Slope
[1] The resistance condition in the table is for reference only.