5322A
Calibration Manual
68
j. Measure the Product resistance in points 2 G
Ω
and higher. Compare the measurement
on the Product with the limits in
2. 100 G
Ω
test:
a. Select the HI
Ω
function on the Product. Push the
Mode
softkey and select
100 G
Ω.
b. Push
O
.
c. Measure the Product resistance. Compare the measurement on the Product with the
limits in
3. OPEN High resistance source test:
a. Connect the 8588A to the low resistance source output terminals LO
Ω
HI and LO. Use a
two-wire connection.
b. Select HI
Ω
function in the Product. Push the
Mode
softkey and select
OPEN
.
c. Push
O
.
d. Measure the Product resistance. The value reads >90 M
Ω
.
4. SHORT high resistance source test:
a. Connect the 8588A to the low resistance source output terminals LO
Ω
HI and LO. Use a
two-wire connection.
b. Select the HI
Ω
function on the Product. Push the
Mode
softkey and select
SHORT
.
c. Push
O
.
d. Measure the Product resistance. The value reads
≤
250
Ω
.
5. Test voltage High resistance source test:
a. Connect the 5522A voltage output terminals to the high resistance source output
terminals HI
Ω
HI and LO. Select the 5522A dc voltage function.
b. Select HI
Ω
function on the Product. Push the
Mode
softkey and select
Resistance
.
c. Push
O
.
d. Set the resistance value for the 5 kV and 1.5 kV option with
e. Set the appropriate voltage on the 8588A.
f. Measure the test voltage on the Product. Do not exceed the limits in
and
for the 5 kV and 1.5 kV option.
6. Short current High resistance source test:
a. Connect the 5522A current output terminals to the high resistance source output HI
Ω
HI and LO terminals. Select the dc current function in the 5522A. Set output current to
10 mA.
b. On the Product, select the HI
Ω
function. Push the
Mode
softkey and select
SHORT
.
c. Push
O
and switch 5522A output terminals on.