5322A
Calibration Manual
88
List of Error Messages
See
for a list of error messages shown by the Product.
Table 34. Error Messages
ID
Error message
Description
-430
Deadlocked.
Remote interface error. A received command generates
too much data to fit in the output buffer and the output
buffer is full. Command execution continues but all data
as lost.
-420
Unterminated.
Remote interface error. The Product was addressed to
talk but a command was not received which sends data
to the output buffer.
-410
Interrupted.
Remote interface error. A received command sends data
to the output buffer, but the output buffer contained
data from a previous command. The output buffer is
cleared when power turns off, or after you execute a
reset command.
-363
Input buffer
overrun
Remote interface error.
-220
Invalid parameter
Remote interface error. An invalid character string was
received. Check to see if the character string was
enclosed in single or double quotes and that the string
contains valid ASCII character.
-140
Character data
Remote interface error.
-120
Numeric data
Remote interface error.
-110
Command header
Remote interface error. A command was received that is
not valid for the Product. The command could be
misspelled.
-103
Invalid separator
Remote interface error. An invalid separator was found
in command string. It might contain a comma instead of
a colon, semicolon, or blank space – or there could be a
blank space instead of comma.
501
Eeprom write.
Eeprom write failed.
502
Eeprom read.
Eeprom read failed.
503
Eeprom error.
Eeprom data lost.
651
Impedance too
high.
Residual impedance is too high (SCAN, COMP). Do not
use SCAN and COMP mode in the Loop/Line impedance
function or connect the Product to the socket with lower
residual impedance.
652
Compensator
overload
Compensator cannot be set. Hardware failure.