Multifunction Electrical Tester Calibrator
Adjustment Procedures
33
8. Connect the voltage output of the 5522A to the ZGND PE and N terminals. See
Figure 24. Ground Resistance Meters Adjustment Connection 2
9. Select
GBR V01 Test Voltage 23 VacL
(Open)
and push the
Select
softkey.
10. Set the output voltage to 100 V dc on the 5522A and switch the output terminals ON.
11. Use the rotary knob or cursor keys to adjust the value to 23.0 V.
12. Repeat step 11 for the next position,
GBR V02 230A acH
with the requested voltage of
230.0 V.
13. Push the
Write
softkey to confirm new adjustment value.
This completes this part of the adjustment.
1
4
7
/
+
STBY
5522A CALIBRATOR
Fluke 5322A Fluke 5522A