5322A
Calibration Manual
74
3. DC/AC multimeter phase test:
a. Connect the 5522A current output terminals to A and COM terminals on the Product.
b. Connect the 5522A voltage output terminals to V and COM terminals on the Product.
c. Set frequency and output voltage as in
.
d. Select the
METER
function, and the
Multimeter
mode on the Product.
e. Select
AC
mode.
f. Switch the 5522A output to ON.
g. Compare the measurement on the Product with the limits in
.
4. DC/AC HIPOT leakage current test:
a. Connect the 5522A current output terminals to A and COM terminals on the Product.
b. Set the frequency and output voltage, see
c. Select the
METER
function and the
HIPOT LC
mode on the Product.
d. Select appropriate
AC
or
DC
mode.
e. Switch the 5522A output to ON.
f. Compare the measurement on the Product with the limits in
.
5. Flash V test:
a. Connect the 5522A voltage output terminals to V and COM terminals on the Product.
b. Select
AC
mode and an output of 1000 V at 55 Hz.
c. Select the
METER
function and the
Flash V
mode on the Product.
d. Switch the 5522A output to ON.
e. The measured voltage on the Product should be within the limits of 991 V to 1009 V.
6. Flash LC test:
a. Connect the 5522A voltage output terminals to the A and COM terminals on the
Product.
b. Select
AC
mode and an output of 2 mA at 55 Hz.
c. Select the
METER
function and the
Flash LC
on the Product.
d. Switch the 5522A output to ON.
e. The measured current on the Product should be within the limits of 1.9945 mA to
2.0055 mA.