5322A
Calibration Manual
86
Table 31. AC/DC Phase Multimeter Test
Nominal
Output
Voltage
Nominal
Output Current
Nominal
Phase Shift
Frequency
(Hz)
Lower Limit
Upper Limit
250 mA AC
100 V
0 °
55
- 0. 1 °
+0.1 °
2.5 A AC
100 V
0 °
55
- 0. 1 °
+0.1 °
20 A AC
100 V
0 °
55
- 0. 1 °
+0.1 °
250 mA AC
100 V
45 °
55
44.9 °
45.1 °
2.5 A AC
100 V
45 °
55
44.9 °
45.1 °
20 A AC
100 V
45 °
55
44.9 °
45.1 °
Table 32. DC/AC HIPOT Leakage Current Test
Nominal Output
Voltage
Frequency
(Hz)
Lower Limit
(A)
Upper Limit
(A)
100
μ
A
-
99.50
μ
A
100.50
μ
A
200
μ
A
-
199.20
μ
A
200.80
μ
A
250
μ
A
-
249.05
μ
A
250.95
μ
A
1 mA
-
0.9965 mA
1.0035 mA
2 mA
-
1.99450 mA
2.00550 mA
2.5 mA
-
2.4935 mA
2.5065 mA
10 mA
-
9.965 mA
10.035 mA
20 mA
-
19.945 mA
20.055 mA
25 mA
-
24.935 mA
25.065 mA
100 mA
-
99.65 mA
100.35 mA
200 mA
-
199.45 mA
200.55 mA
250 mA
-
249.35 mA
250.65 mA
100
μ
A
50
99.50
μ
A
100.50
μ
A
200
μ
A
50
199.20
μ
A
200.80
μ
A
250
μ
A
50
249.05
μ
A
250.95
μ
A
250
μ
A
60
249.05
μ
A
250.95
μ
A
250
μ
A
400
249.05
μ
A
250.95
μ
A
1 mA
50
0.9965 mA
1.0035 mA
2 mA
50
1.99450 mA
2.00550 mA
2.5 mA
50
2.4935 mA
2.5065 mA
2.5 mA
60
2.4935 mA
2.5065 mA