5322A
Calibration Manual
76
3. 10 kV divider dc voltage test:
a. Connect the 10 kV divider rear panel output terminals to the V and COM input terminals
on the Product. Use the 5 kV test lead from 5322A accessory on the high voltage side.
Connect the grounding post on the Resistance multiplier to Product rear panel
grounding post.
b. Select the
METER
function on the Product.
c. Push the
Input
softkey to activate the HV input terminal.
d. Select
With 10 kV probe
.
e. Push the
AC/DC
softkey to set the mode to
DC
.
f. Set 4500 V dc on the 19052. Switch the 19052 output terminals ON.
g. The Product should read within the limits of 4481.5V to 4518.5 V.
4. 10 kV divider ac voltage test:
a. Select the
METER
function on the Product.
b. Push the
Input
softkey to activate the HV input terminal.
c. Select
With 10 kV probe.
d. Push the
AC/DC
softkey to set the mode to
AC
.
e. Set 4500 V ac on the 19052. Switch the 19052 output terminals ON.
f. The Product should read within the limits of 4472.5V to 4527.5 V.