Multifunction Electrical Tester Calibrator
Adjustment Procedures
61
This completes this part of the adjustment.
HIPOT Ripple Adjustment
HIPOT Ripple adjustment enables adjustment of the meters used for the ripple coefficient
measurement. Adjustment is done on the ac signal.
To make this adjustment, see
1. Connect the 5522A to the Product output terminals.
2. Push the
Setup
softkey on the Product.
3. Highlight
Calibration
and push the
Select
softkey.
4. Highlight
HIPOT Ripple
and push the
Select
softkey. The list of adjustment points shows,
Table 16. HIPOT Ripple Calibration Points
5. Select the first position with nominal value, 10 V. The Product shows the actual measured
voltage.
6. Set the Product function to ac voltage, 10 V at 55 Hz.
7. On the 5522A, push the ON button and the Product reads approximately 10 V.
8. Push the
Write
softkey to save the new value for the 10 V voltage range. To leave the
original value, push the
Exit
softkey. The Product ignores the measured current and returns
to list of current ranges.
9. Go through and make new adjustments for each of the ranges in the list.
10. When the calibration process concludes, push the
Exit
softkey until the Low resistance
source function shows.
This completes this part of the adjustment.
Nominal Current
Range
Position
Adjustment Point
Frequency
Required
Accuracy of
Adjustment
10 V ac
Range 10 V 55 Hz
high (10 V)
10 V
55 Hz
0.1 %
100 V ac
Range 100 V 55
Hz high (100 V)
100 V
55 Hz
0.1 %
1000 V ac
Range 1000 V 55
Hz high (950 V)
950 V
55 Hz
0.1 %