5322A
Calibration Manual
50
AC Ranges 10 V, 100 V, 1000 V Adjustment
To adjust these ac ranges:
1. Connect the 5522A or similar to the Product output terminals, see
Figure 35. AC Ranges 10 V, 100 V, and 1000 V Adjustment Connections
2. Push the
Setup
softkey on the Product.
3. Select
Calibration
and push the
Select
softkey.
4. From the calibration menu, highlight
AC Multimeter
and push the
Select
softkey. The list of
adjustment points shows.
5. Highlight the first range with the nominal value,
10 V 55Hz low (1V)
and push the
Select
softkey.
6. Set 1 V ac 55 Hz on the 5522A. Switch the output terminals ON. The Product reads
approximately the nominal value.
7. Push the
Write
softkey to save new value for the 10 V range. To leave the original value,
push the
Exit
softkey. The Product ignores the measured current and returns to list of
current ranges.
8. Do the same adjustment for all ac voltage ranges, 10 V, 100 V, and 1000 V.
This completes this part of the adjustment.
1
4
7
/
+
STBY
5522A CALIBRATOR
Fluke 5322A Fluke 5522A