5322A
Calibration Manual
82
Test Voltage
Frequency
(Hz)
Lower Limit
Upper Limit
200 V
-
199.62 V
200.38 V
300 V
-
299.52 V
300.48 V
400 V
-
399.42 V
400.58 V
500 V
-
499.32 V
500.68 V
600 V
-
599.22 V
600.78 V
60 mV
50
58.7 mV
61.3 mV
200 mV
50
198 mV
202 mV
200 mV
60
198 mV
202 mV
200 mV
400
198 mV
202 mV
270 mV
50
267.65 mV
272.35 mV
600 mV
50
595.2 mV
604.8 mV
2 V
50
1.991 V
2.009 V
2 V
60
1.991 V
2.009 V
2 V
400
1.991 V
2.009 V
2.7 V
50
2.689 V
2.711 V
6 V
50
5.985 V
6.015 V
20 V
50
19.971 V
20.029 V
20 V
60
19.971 V
20.029 V
20 V
400
19.971 V
20.029 V
80 V
50
19.971 V
20.029 V
80 V
60
79.89 V
80.11 V
80 V
400
79.89 V
80.11 V
110 V
50
109.80 V
110.2 V
150 V
50
149.76 V
150.24 V
200 V
50
199.71 V
200.29 V
250 V
50
249.66 V
250.34 V
280 V
50
279.63 V
280.37 V
280 V
60
279.63 V
280.37 V
280 V
400
279.63 V
280.37 V
500 V
50
499.32 V
500.68 V
500 V
60
499.32 V
500.68 V
500.00
400 499.32
V
500.68
V
Table 25. DC/AC Voltage Calibrator (cont.)