7.2.2 Current output 1 / 2
With single-circuit measurement the
configuration is always as follows:
Current output 1: pH / redox
Current output 2: temperature
With two-circuit measurement (see 7.1.1):
Current output 1: pH / redox (channel 1)
Current output 2: pH / redox (channel 2)
or
temperature
or
diff. (channel 1 – channel 2)
Function
Selection
Default
Current loop alarm
(Broken line activates fault contact,
error message 15 / 16)
active,
inactive
inactive
Error current
off
(no error current)
min. current
1)
max. current
2)
off
Current output 1
Measuring current range
0 ... 20 mA
4 ... 20 mA
4 ... 20 mA
Damping
1.0 ... 20.0 mA/s
20.0 mA/s
Measuring range lower limit
(measured value at 0 / 4 mA)
pH:
–2.00 ... +16.00 pH
Redox mV:
–1500 ... +1500 mV
Redox %:
–3000 ... +3000 %
0.00 pH
–500 mV
0 %
Measuring range upper limit
(measured value at 20 mA)
3)
pH:
–2.00 ... +16.00 pH
Redox mV:
–1500 ... +1500 mV
Redox %:
–3000 ... +3000 %
12.00 pH
+500 mV
100 %
Current output 2
Measuring current range
0 ... 20 mA
4 ... 20 mA
4 ... 20 mA
Damping
1.0 ... 20.0 mA/s
20.0 mA/s
Measuring range lower limit
(measured value at 0 / 4 mA)
Temperature:
–20.0 ... +150.0 °C
pH 2
4)
:
–2.00 ... +16.00 pH
Redox mV 2
4)
: –1500 ... +1500 mV
Redox % 2
4)
:
–3000 ... +3000 %
∆
pH
4)
:
0.00 ... 18.00 pH
∆
mV
4)
:
0 ... 3000 mV
∆
%
4)
:
0 ... 6000 %
0.0 °C
0.00 pH
–500 mV
0 %
0.00 pH
0 mV
0 %
Measuring range upper limit
(measured value at 20 mA)
3)
Temperature:
–20.0 ... +150.0 °C
pH
4)
:
–2.00 ... +16.00 pH
Redox abs.
4)
:
–1500 ... +1500 mV
Redox rel.
4)
:
–3000 ... +3000 %
∆
pH
4)
:
0.00 ... 18.00 pH
∆
mV
4)
:
0 ... 3000 mV
∆
%
4)
:
0 ... 6000 %
150.0 °C
12.00 pH
+500 mV
100 %
18.00 pH
1800 mV
1800 %
1)
0.00 mA for measuring current range 0 ... 20 mA;
2.40 mA for measuring current range 4 ... 20 mA
2)
22.00 mA
3)
Min. difference between upper and lower limit:
∆
1.8 pH,
∆
400 mV,
∆
10% bzw.
∆
17.1 °C
4)
Only with two-channel measurement
Instrument configuration
mycom CPM 152
36