Function
Selection
Default
If »parameterize« is selected
P controller
Control gain
K
p
: 0.10 ... 20.00
1.00
Controller output
Min. controller output: 0 ... 50 %
Max. controller output: 50 ... 100 %
0 %
100 %
Control effect
Direct, inverse
inverse
PI controller
Control gain,
integral action time
K
p
: 0.10 ... 20.00
T
n
: 0.1 ... 999.9 min
1.00
10.0 min
Controller output
Min. controller output: 0 ... 50 %
Max. controller output: 50 ... 100 %
0 %
100 %
Control effect
Direct, inverse
inverse
PD controller
Control gain,
derivative action time
K
p
: 0.10 ... 20.00
T
v
: 0.1 ... 999.9 min
1.00
1.0 min
Controller output
Min. controller output: 0 ... 50 %
Max. controller output: 50 ... 100 %
0 %
100 %
Control effect
Direct, inverse
inverse
PID controller
Control gain,
integral action time,,
derivative action time
K
p
: 0.10 ... 20.00
T
n
: 0.1 ... 999.9 min
T
v
: 0.1 ... 999.9 min
1.00
10.0 min
1.0 min
Controller output
Min. controller output: 0 ... 50 %
(corresponds to basic
load metering)
Max. controller output: 50 ... 100 %
0 %
100 %
Control effect
Direct, inverse
inverse
If »operating mode« is selected
Changing operation mode
Controller automatic
Controller manual
manual
Controller output (if »manual«
is selected)
0 ... 100 %
0 %
Setpoint (if »automatic« is chosen)
–2.00 ... +16.00 pH
4.00 pH
↑↓
MAN_OP.EPS
Bild 8.7
The display in operating
mode »Manual«
Symbol for manual mode
Controller configuration
mycom CPM 152
52