6.6
Access codes
Functions can be protected by four-digit
access codes to make sure no unintentional
or undesirable changes are made to the
configuration and calibration data of the
measuring transmitter.
The release status of the functions
is
divided up into
operator and advanced
level.
Menus that have not been released are
not displayed. If there is no input or the
input is incorrect, the code input menu can
only be left via the MEAS key.
The access codes are set in the menu
configuration / commissioning / system set-up.
(see chapter 7.1.2).
Note:
The instrument is supplied unlocked.
Accessible
without code
:
• Measurement displays
• Fault list
• Info list
• Logbook
• Calibration statistics
Accessible
with operator code
:
• Short-cut to relays
• Calibration type selection
• Calibration functions
• Temperature compensation
• Cleaning functions
Accessible
with advanced code
:
• All menus and functions
Caution:
If the codes get lost, unlocking is
possible with the default code
“7156”
and a new code can be set
on the instrument data / system
configuration menu.
↑→
MPDIS07E.EPS
Fig. 6.4
Code enquiry