Display
Description
Calibration is being carried out now. The
instrument is waiting until the pH
measurement has stabilised and will then
store the measured value and move on to
the calibration information.
If the measured value is not stable after 300
seconds a display appears with the option of
discontinuing the calibration operation.
Press the E key to stop the calibration here
and to return to the “Type of calibration”
menu.
Display of the buffers identified.
Calibration information.
If the electrode zero point and slope figures
are outside the tolerance limits you will be
informed about this on the display. If the pH
values of the buffer solutions used are too
close together, you will be informed about
this on the display. Repeat the calibration
operation with suitable buffers.
Display with plausible calibration values.
Final menu. You can:
●
store the calibration
●
repeat the calibration
●
discard the calibration and
return to measuring mode
Make your choice and press E.