8.3.2
Controller (equipment with plug-in module FCYK)
– 3 additional contacts
Function
Selection
Default
If a controller is set
three configuration groups
can be selected.
Configuration
Parameterize
Operation mode
configuration
If »configuration« is selected
Selection of type of controller
None,
P-, PI-, PD, PID controller,
Three-point step contr. (PD, PDT 1),
Neutralisation controller,
2-range P controller
(the last four options can be
selected by scrolling the list)
none
For neutralisation controller:
controller type
P, PI, PD, PID controller
P controller
Selection of output
Control contact 1,
control contact 2 (not fpr P, PI, PD,
PID controller),
alarm 1
1)
control
cont. 1
If »Control contact 1« or »Control contact 2« is selected
Controller characteristics
Pulse-length output
Pulse-frequency output
3-point step + (output for positive
actuating signals),
3-point step – (output for negative
actuating signals)
(depends on
controller
type)
Pulse cycle time
(for »Pulse-length output«
or »3-point step +«)
0.5 ... 99.9 s
10.0 s
Min. cycle time (for »Pulse-length
output« or »3-point step +«)
0.1 ... 5.0 s
0.3 s
Maximum pulse frequency
(if »Pulse-frequency output«
is selected)
60 ...180 1/min
120 1/min
Type of contact
N.C., N.O.
N. O.
If »Alarm 1« is selected
Alarm threshold
(related to set point)
0.10 ... 4.00 pH
1.00 pH
Alarm delay
0 ... 6000 s
0 s
1)
Exceeded setpoint result in error number E 067 and can be signalled at the
»maintenance« contact.
Continued on the next page
Function
Selection
Default
If »Operation mode« is selected
Selection of limit contact
Limit contact 1, limit contact 2
limit cont. 1
Changing operating mode
Auto contact 1 / 2
Manual contact 1 / 2
manual
contact
Manual operation of limit contactor
(If »Manual contact 1 / 2« is selected)
Off, on
off
Controller configuration
mycom CPM 152
54
Hauser