13
Appendix
13.1
Technical data
13.1.1
General technical data
General specifications
Manufacturer
Hauser
Product designation
Mycom CPM 152
pH measurement
Measuring range
pH –2.00 ... +16.00
Measured value resolution
pH 0.01
Deviation of indication
1
, measured value
max. 0.2 % of measuring range
Reproducibility
1
max. 0.1 % of measuring range
Zero-point shift range
pH –2.00 ... +16.00
Automatic temperature compensation range
–20 ... +150 °C
Reference temperature
25 °C
Slope adjustment
5 ... 99 mV/pH
pH signal input
Input resistance at nominal operating
conditions
> 1 × 10
12
Ω
Input current at nominal operating conditions < 1.6 × 10
–12
A
pH signal output
Current range (active)
0 / 4 ... 20 mA
Measurement deviation
1
max. 0.2 % of current range maximum
Load
max. 600
Ω
(Ex: max. 500
Ω
)
Transmission range
adjustable,
∆
1.8 ...
∆
18 pH
Redox measurement
Measuring range
–1500 ... +1500 mV / –3000 ... + 3000 %
Measured value resolution
1 mV / 1 %
Deviation of indication
1
, measured value
max. 0.2 % of measuring range
Reproducibility
1
max. 0.1 % of measuring range
Assignment for % display
adjustable,
∆
for 100 % = 150 ... 2000 mV
Electrode offset
±200 mV
Redox signal input
Input resistance at nominal operating
conditions
> 1 × 10
12
Ω
Input current at nominal operating conditions < 1.6 × 10
–12
A
Redox signal output
Current range (active)
0 / 4 ... 20 mA
Measurement deviation
1
max. 0.2 % of current range maximum
Load
max. 600
Ω
(Ex: max. 500
Ω
)
Transmission range
adjustable,
∆
300 ...
∆
3000 mV
mycom CPM 152