Set-up guide / checklist
Question
see
chapter
Possible choices
Factory settings
User settings
Language
7.1.4
Language version A:
Deutsch, English, Français, Italiano
Language version C:
Deutsch, English, Français, Nederlands,
Japanese
English
Contrast of the
LCD display
7.1.4
Setting as required
average
Date
6.4
Enter the date
current date
Time
6.4
Enter the time
Central European
Time,
(not summer time)
Measurement
parameter
7.1.1
pH,
redox absolute (mV),
redox relative (%)
pH
Type of measurement
(only with two-circuitl
measurement of pH)
7.1.1
Single circuit measurement channel 1,
Single circuit measurement channel 2,
redundancy measurement
Redundancy
measurement
Type of electrode 1
7.1.1
Glass electrode pH 7.0;
glass electrode pH 4.6;
antimony
Glass electr. pH 7.0
Type of electrode 2
(only with two-circuit
measurement of pH)
7.1.1
Glass electrode pH 7.0;
glass electrode pH 4.6;
antimony
Glass electr. pH 7.0
Potential matching
7.1.1
without PM (unsymmetrical)
with PM (symmetrical high-impedance)
with PM
Unit for temperature
measurement
7.3.2
Celsius [°C],
Fahrenheit [°F],
Kelvin [K]
Celsius [°C]
Temperature
compensation
(only pH meas.)
7.3.1
ATC (automatic) channel 1 / channel 1+2,
MTC (manual) Pt 100,
MTC (manual)
ATC
Temperature
measurement
(only redox meas.)
7.3.2
on,
off
On
Contact function
7.1.3
basic configuration:
Fault + maintenance,
Fault + contact
Fault +
maintenance
7.1.3
if equipped with 3 additional contacts:
3 alarms, 2 contacts
fault, Chemoclean, contacts
3 alarms, Chemoclean
alarms, Chem holder
3 alarms, 2 contacts
Alarm contact
7.1.3
N.C. contact,
N.O. contact
N.O. contact
Fault contact
7.1.3
Pulsed contact,
continuous contact
Continuous contact
End of set-up
End,
cancel (start again)
End
pm152e05.chp
mycom CPM 152
First Start-up
Hauser
21