9.3.4
Calibration relative (%)
A sample of the medium is filled in two
containers for calibration purposes. The
contents of the first container are detoxified
and serve as buffer 1.
The contents of the second container are left
unchanged and serve as buffer 2.
Display
Description
Wait until the redox measurement has
stabilised.
Press the E key to store the measurement.
Enter the relevant redox value of buffer 1
(detoxified sample) in percent.
The redox value measured with the old
calibration data is displayed. Wait until the
measurement has stabilised.
Press the E key to store the measurement.
Enter the relative redox value of buffer 2 in
per cent.
Continued on the next page
1
E
↑↓→
2
E
↑↓→
Detoxified
sample
Unchanged
sample
Calibration
mycom CPM 152
66
Hauser