SCS controlled cleaning
Additionally to one of the three cleaning
programs, the cleaning process can be
actuated by the SCS fault messages E030
and E031 if the »SCS« electrode monitoring
system is switched on (see Chapter 7.4).
Note:
•
The total SCS electrode monitoring
is only possible in the case of
symmetrical measurement
with PAL.
•
If the SCS fault cannot be
eliminated by two cleaning
operations, the SCS-controlled
cleaning facility is disabled.
7.6.4 Parameterize Chemoclean
Function
Selection
Default
Switching cleaning on / off, set
parameters
Chemoclean on,
Chemoclean off,
Set-up
Chemoclean
off
If »set-up« is chosen
Type of cleaning programme
interval cleaning,
interval measurement
week programme
interval
cleaning
Interval cleaning
Cleaning cycle: interval time
(time between two cleanings)
6 min ... 99 h 59 min 59 s
8 h
Continuation see »Settings for all cleaning programmes«
Interval measurement
Dwell time
(electrode not in the medium)
0 s ... 99 h 59 min 59 s
0 min
Measurement time
(electrode in the medium)
1 min ... 99 h 59 min 59 s
1 min
Continuation see »Settings for all cleaning programmes«
Week programme
Week programme parameters
Set-up, cancel
set-up
Week programme
Number of starts (0 ... 12)
chosen individually for every day of
the week (Mon ... Sun)
0
Day programme
Individual starting time for each
start number
0 h
Next day
Editing the next day of the week,
discontinuation
next day
Continuation see »Settings for all cleaning programmes«
Instrument configuration
mycom CPM 152
46