11.4.3 Factory reset
Caution:
If calibration data are reset,
the measuring system must
be re-calibrated.
11.4.4 Instrument check
Function
Selection
Default
(Reset options)
Cancel (no reset),
set config (store modified module assignment),
Only configuration data,
Only calibration data,
All data,
Service data
set config
After pressing the E key the hardware equipment is
checked and the modified slot assignment is stored.
Only configuration data
reset
All configuration data are reset to the factory setting if
confirmed by pressing the E key.
Caution:
All previous instrument configuration data
are lost!
Only calibration data reset
All calibration data are reset to the factory setting if
confirmed by pressing the E key.
Caution:
All previous calibration data are lost!
All data reset
Configuration and calibration data are reset to the factory
setting if confirmed by pressing the E key.
Caution:
All previous configuration and
calibration data are lost!
Service data,
logbook,
reset counter,
message logbook
Functions for authorized service personnel only.
Service Code necessary.
Function
Description
Type of test
Selection: keyboard, display, RAM, EPROM, EEPROM
Keyboard
Graphic presentation of the keyboard layout.
Press all the keys in succession. If the key is working an
acknowledgement appears in the relevant field of the
display. You return to selection type of test afterwards by
pressing the E key.
Display
A chessboard pattern and its inversion are shown
alternately. Check the display to see if any dots are
missing. You return to selection type of test by pressing
the E key.
RAM
Internal test. At the end of the testing time the result is
displayed. You return to “Selection type of test” by
pressing the E key.
EPROM
EEPROM
pm152e11.chp
mycom CPM 152
Diagnostics