8.2.2
Controller (standard equipment)
Continued on the next page
Function
Selection
Default
If limit contactor 1 »on«
three configuration groups
can be selected.
Limit value configuration
alarm configuration
operating mode
Limit value
configuration
If »Limit value configuration« is selected
Switching limit contact 1 on / off
on, off
off
Limit value
–2.00 ... +16.00 pH
4.00 pH
Hysteresis
0.10 ... 1.00 pH
0.50 pH
Alignment of the limit
min. function
max. function
min. function
Pickup delay
0 ... 7200 s
0 s
Dropout delay
0 ... 7200 s
0 s
Contact type
N.C., N.O.
N.O.
If »Alarm configuration« is selected
Alarm threshold (related to limit value)
0.10 ... 18.00 pH
1.00 pH
Alarm delay
0 ... 6000 s
0 s
If »Operating mode« is selected
Changing operating mode
automatic limit contact 1,
manual limit contact 1
manual
Function
Selection
Default
If any controller is set, three
configuration groups can be
selected.
configuration
parameterize
operation mode
configuration
If »configuration« is selected
Control characteristics
P, PI, PD, PID, none
none
Selection of the output
control contact 1
alarm 1
1)
contact 1
If »Control contact 1« is selected
Signal outputs
pulse length,
pulse frequency
pulse length
Pulse cycle time (if »Pulse-length
controller« is selected)
0.5 ... 99.9 s
10.0 s
Min. cycle time (if »Pulse-length
controller« is selected)
0.1 ... 5.0 s
0.3 s
Max. pulse frequency (if »Pulse
frequency« is selected)
60 ... 180 1/min
120 1/min
Type of contact
N.C., N.O. relay
N.O.
If »Alarm 1« is selected
Alarm threshold (related to set point)
0.10 ... 18.00 pH
1.00 pH
Alarm delay
0 ... 6000 s
0 s
1)
Exceeded setpoint result in error number E 067 and can be signalled at the
»maintenance« contact.
pm152e08.chp
mycom CPM 152
Controller configuration
Hauser
51