4.5.3
Connection accessories
Extension cables
Junction box VBA
This junction box is required for line lengths of
more than 20 m between the assembly and
the transmitter. It has 4 Pg glands for cable
entry and 10 insulated high-impedance
screwed terminals for connection of individual
wires.
Dimensions: 120 x 120 x 55 mm (L x W x H);
Material: plastic; protection class: IP 65;
Order no.: 50005276
Caution:
The desiccant bag has to be
checked and if necessary replaced
at certain intervals (determined by
the ambient conditions), in order to
avoid measuring faults due to
moisture bridges on the
pH connection line.
4.6
Removal, packaging and disposal
Repackaging
The instrument must be protected against
shock and moisture when it is packaged for
later reuse. The original packaging provides
the most effective protection.
Disposal
Note:
Electronics scrap is hazardous
waste! Make sure you observe the
local regulations for its disposal!
CYK 71
DMK
CYK-DMK.CDR
7.2
10.2
Fig. 4.25
Extension cables:
CYK 71 (for CPK 1, 7, 9)
and DMK (for CPK 6)
vba-dose.tif
18
0
12
0
90
110
Height 55
Fig. 4.26
Dimensions
junction box VBA
pm152e04.chp
mycom CPM 152
Installation
Hauser
19