11.1.3 Error table
Failure
No.
Display
Remedy
E001
ERROR occured on internal data
(red DIAG-LED endures even after
the error has become inactive)
Return the instrument to your
Hauser sales agency to be
repaired or request service assistance.
E002
Data ERROR in EEPROM
E003
Invalid module configuration
Check slot configuration
E004
Module configuration changed
To set new slot configuration select
“factory reset/hardware config” in
service menu (see 10.4.3)
E005
Unknown module card-ID
Check slot configuration
E006
Checksum error in EEPROM
Correct the checksum with the function
»Service«. If successless, change
EEPROM.
E007
Transmitter defective
When temperature measurement is
correct, check sensor and connections.
When temperature measurement is not
correct, change the transmitter.
E008
SCS glass broken
Check the pH electrode for glass
breakage; check electrode plug-in head
for moisture and dry it if necessary;
check medium temperature. (Maximum
cable length for SCS is 20 m).
E009
SCS glass broken, electrode 2
E010
Temperature sensor defective
Check temperature measurement and
connections; check instrument and
measuring cable with temperature
simulator if necessary
E011
Temperature sensor 2 defective
E013
Collecting error assembly
Check retractable assembly
(compressed air, mechanical blockage)
E015
Current loop 1 open
Check connections lines and any
connected instruments
E016
Current loop 2 open
E019
Delta threshold overflow
Check delta window settings (see 7.4)
E020
Voltage input range under limit
Repeat calibration and replace buffer
solutions; if necessary replace the
electrode and check instrument and
measuring cable with a simulator
E021
Voltage input range 2 under limit
E022
Voltage input range over limit
E023
Voltage input range 2 over limit
Maintenance
E030
SCS fault reference electrode
Check the reference electrode for glass
breakage and dirt; clean the reference
electrode; check the temperature of the
medium
E031
SCS fault reference electrode 2
E032
Out of the set slope range
Repeat calibration and replace buffer
solutions; if necessary replace the
electrode and check instrument and
measuring cable with simulator
E033
Out of the set asymmetry range
E034
Out of the set offset range