Note:
If there are discrepancies in the
operating mode »week programme«
(minimum interval between the
operations: 0.1 h or 6 min), the
following error message appears:
»Interval < 6 min«.
7.6.5
Action when the supply systems fail (mains power / compressed air)
Action when a retractable assembly is
connected:
If there is a retractable assembly connected,
it always stays in its current position when a
failure occurs in the compressed air or power
supply system or when there is a fault in the
Mycom CPM 152 measuring transmitter
(faults E001 - E023).
Failure during a cleaning cycle:
The cleaning operation is repeated when the
supply system is working again.
This does not apply to cleaning operations
that have been started manually on the
»short-cut to relays« menu.
Failure during interval cleaning or
measurement in measuring mode:
If it has not been possible to carry out
cleaning operations during a supply failure,
the last cleaning operation is carried out when
the supply system is working again.
Apart from this, the cleaning program is
continued as is no failure has occurred.
If the failure occurred while the program was
in measuring mode, one minute is measured
before any further action is taken, in order to
make sure that a stable reading is obtained.
Failure during the week programme in
measuring mode:
Cleaning operations during a failure are
disregarded and are not carried out
afterwards.
Function
Selection
Default
Settings for all cleaning programmes
Pre-rinsing time
0 ... 999 s
10 s
Cleaning time
0 ... 999 s
5 s
Post-rinsing time
0 ... 999 s
10 s
Repeat rate
0 ... 5 times
0 times
Number of cleaning cycles with
cleaning agent
0 ... 9 times
0 times
Hold during Chemoclean
On, off
on
Hold continuation time (to allow the
measurement to settle)
0 ... 999 s
10 s
SCS cleaning
(starts if error code E030/E031
occurs while SCS is active)
On, off
off
Total interval
(if »Measurement at intervals« is
selected)
No selection, frequency and total
duration of the cleaning cycle plus
the dwell and measurement time
are displayed
Total cleaning time (if »Cleaning at
intervals«, »Weekly programme«
are selected)
No selection, the total duration of a
cleaning cycle is displayed
pm152e07.chp
mycom CPM 152
Instrument configuration
Hauser
47