6.7
The “short-cut to relays” menu
The short-cut to relays menu gives you direct
access to the main functions without having to
go through the whole of the configuration
menu.
Note:
Functions which are not available or
not active are not shown in the menu.
Short-cut to relays menu
Function
Selection
Hold
1)
Hold on / hold off
Limit contact operation mode
Limit contact automatic
Limit contact manual
If manual mode selected
Contact on / off
Note:
•
The outputs can be set in spite of a
possibly activated hold function
•
The following symbol for
manual operation appears
in the measured value display:
Controller operation mode
Controller automatic
Controller manual
If manual mode selected
Manual modification of the actuating variable
Note:
•
When a switch is made to manual
operation, the present setting is used as
the starting point for the manual setting
(smooth switching)
•
The following symbol for
manual operation appears
in the measured value display:
Set values
Depending on the configuration
(see chapters 8.2/8.3):
Input of limit value 1/2 or controller set value
Chemoclean
Automatic on
Automatic off
Start cleaning
Holder
2)
Service
(if holder in
S Clean
measuring position)
Cleaning
(if holder in
Finish service
service position)
1)
The hold function does not take effect on activated Chemoclean cleaning cycles.
2)
After “Service” is selected the assembly is brought into service position.
Automatic and external start of cleaning cycles will be disabled.
The assembly can be brought back into measuring position only with “Finish service”.
→
Short-cut to relays
PARAM
Operation
mycom CPM 152
28
Hauser