6
Troubleshooting
6.1
Finding faults with the keypad
WARNING!
ELECTRIC SHOCK
Install all protective equipment ground covers and safety ground leads after troubleshooting.
Failure to install covers and ground leads could result in serious injury or death.
If a fault occurs, the control unit display switches from its current mode of operation to the
DIAGNOSTICS
mode and displays the current fault condition. Refer to
further information on the fault conditions. If the fault clears, the display stays in
DIAGNOSTICS
mode and displays
All Clear
.
Enter
DIAGNOSTICS
mode from the instrument's keypad; you can do this at any time
without interrupting or disturbing the analog outputs of the equipment. As an initial guide
to equipment performance, typical values for instrument operation are given below.
D2 min
D2 max
D1 min
D1 max
Set Cal min
Set Cal max
Rosemount
™
CCO 5500
3,000
20,000
3,000
20,000
2,000
2,900
where:
Modulation frequency = 30 to 45 Hz
Mark/space ratio = 0.9 to 1.1
Saturation count = 50 maximum for all analyzers
If the values are outside of the above ranges, the Data valid LED extinguishes, and the fault
condition is displayed.
6.1.1
Data valid LED out
If one or more fault conditions occur, the data valid LED on the front panel extinguishes,
the data valid relay operates, and the instrument automatically enters the
DIAGNOSTICS
mode to display the fault condition.
Fault Conditions
Table 6-1:
Fault condition
Possible causes
Detector signal saturated (Sat # over 50)
Incorrect detector gain adjustment.
Troubleshooting
Reference Manual
91
Summary of Contents for Rosemount CCO 5500
Page 4: ......
Page 8: ...Contents iv Rosemount CCO 5500 ...
Page 28: ...System Wiring Diagram Figure 2 7 Install 20 Rosemount CCO 5500 ...
Page 32: ...Install 24 Rosemount CCO 5500 ...
Page 88: ...Operation 80 Rosemount CCO 5500 ...
Page 98: ...Maintenance 90 Rosemount CCO 5500 ...
Page 108: ...Troubleshooting 100 Rosemount CCO 5500 ...
Page 110: ...Spare parts 102 Rosemount CCO 5500 ...
Page 122: ...Return equipment to the factory 114 Rosemount CCO 5500 ...
Page 123: ...Return equipment to the factory Reference Manual 115 ...