You can reset the average values that are currently held in the four averaging stacks using
this option; this ereases the current average that is held in all of the averaging stacks.
Select this option by pressing ENTER and using the arrow keys. The instrument requests
confirmation before the averages are reset.
Note
If you select this option, all the data in the averaging stacks is reset, and the data for as much as 30
days is lost.
4.10.7
Calibrate
From this option, you may display the two detector levels and conduct a basic calibration.
While in this mode, the gas cell is not moved; this gives an immediate response for setting
up the detector levels. A Cal Factor that is calculated during a calibration routine sets the
basic calibration of the instrument. Press ENTER while this is displayed to see the following
options:
1.
Set Detectors
2.
Span Adjust
3.
Calibrate
Set detectors
You can display both the D1 and D2 levels; you can also display saturation counts. To give
an immediate response to any alterations that are required, the filters and gas cells are not
moved during this operation.
for a discussion of the detector level and saturation count.
Span adjust
Note
Rosemount
™
initially sets the span factor at the factory; do not adjust it unless the instrument
sensitivity is suspected. In any case, Rosemount recommends that you record the orginal value
before making adjustments.
Operation
72
Rosemount CCO 5500
Summary of Contents for Rosemount CCO 5500
Page 4: ......
Page 8: ...Contents iv Rosemount CCO 5500 ...
Page 28: ...System Wiring Diagram Figure 2 7 Install 20 Rosemount CCO 5500 ...
Page 32: ...Install 24 Rosemount CCO 5500 ...
Page 88: ...Operation 80 Rosemount CCO 5500 ...
Page 98: ...Maintenance 90 Rosemount CCO 5500 ...
Page 108: ...Troubleshooting 100 Rosemount CCO 5500 ...
Page 110: ...Spare parts 102 Rosemount CCO 5500 ...
Page 122: ...Return equipment to the factory 114 Rosemount CCO 5500 ...
Page 123: ...Return equipment to the factory Reference Manual 115 ...