If the code is correct after you press ENTER on the last digit, the sequence is
continued. If it is not correct, the instrument returns to
OPERATING
mode. Refer to
Note
Rosemount
™
sets the code number to
0000
at the factory; you should change it in
SET UP
mode.
4.10.2
Set averages
The instrument calculates four separate averages. These are defined in units of seconds,
minutes, hours, and days. You can use any of the four averaging stacks to provide the
instrument's analog output. You can set each averaging time within predefined limits.
1.
Press ENTER when this display is shown.
The display now shows one of the averages.
2.
Use the arrow keys to select the average time that you need to change and press
ENTER to change it.
You can now change the value using the arrow keys and confirm it by pressing
ENTER.
3.
Set the seconds averaging stack to the required value.
This is limited to within 10 to 60 seconds in 10-second intervals.
4.
Set the minutes averaging stack to the required value.
This is limited to within 1 to 60 minutes in 1-minute intervals.
5.
Set the hours averaging stack to the required value.
This is limited to within 1 to 24 hours in 1-hour intervals.
6.
Set the days averaging stack to the required value.
Operation
62
Rosemount CCO 5500
Summary of Contents for Rosemount CCO 5500
Page 4: ......
Page 8: ...Contents iv Rosemount CCO 5500 ...
Page 28: ...System Wiring Diagram Figure 2 7 Install 20 Rosemount CCO 5500 ...
Page 32: ...Install 24 Rosemount CCO 5500 ...
Page 88: ...Operation 80 Rosemount CCO 5500 ...
Page 98: ...Maintenance 90 Rosemount CCO 5500 ...
Page 108: ...Troubleshooting 100 Rosemount CCO 5500 ...
Page 110: ...Spare parts 102 Rosemount CCO 5500 ...
Page 122: ...Return equipment to the factory 114 Rosemount CCO 5500 ...
Page 123: ...Return equipment to the factory Reference Manual 115 ...