Chopper Frequency Trim Pot
Figure 3-7:
A. Trim pot VR1
3.
To increase the chopper frequency, turn the trim pot counterclockwise.
3.7
Set up mode
To prevent any unauthorized changes, you must enter a four digit security code to enter
the set up mode.
You must set operating parameters in the instrument for proper analyzer operation. All
operating parametrs are set within the control unit using the
SET UP
mode. In the
SET UP
mode, parameters are held in non-volatile memory and retained in the event of a power
loss.
Even if the measured data is not going to be normalized, you must set the normalizing
parameters to ensure proper analyzer operation.
Note
When you select
SET UP
mode, the instrument suspends motor operations, and the Data Valid LED
goes dim. If no key is pressed within five seconds after selecting
SET UP
mode, the instrument control
reverts to
OPERATING
mode.
lists all parameters in full. Basic details are given here for configuration purposes.
To aid configuration and to record any subsequent changes to the operating parameters,
Rosemount recommends completing to provide a record of the instrument setup.
Configuration and startup
Reference Manual
35
Summary of Contents for Rosemount CCO 5500
Page 4: ......
Page 8: ...Contents iv Rosemount CCO 5500 ...
Page 28: ...System Wiring Diagram Figure 2 7 Install 20 Rosemount CCO 5500 ...
Page 32: ...Install 24 Rosemount CCO 5500 ...
Page 88: ...Operation 80 Rosemount CCO 5500 ...
Page 98: ...Maintenance 90 Rosemount CCO 5500 ...
Page 108: ...Troubleshooting 100 Rosemount CCO 5500 ...
Page 110: ...Spare parts 102 Rosemount CCO 5500 ...
Page 122: ...Return equipment to the factory 114 Rosemount CCO 5500 ...
Page 123: ...Return equipment to the factory Reference Manual 115 ...