3.
To ensure that the detector signal is not saturating, observe the saturation count
signal displayed next to the detector levels. If a SAT # of more than 0 is displayed,
turn the trim pot slightly to reduce the gain. Reduce the gain until a SAT # of 0 is
displayed.
4.
If saturation is indicated with the trim pot turned fully clockwise, reduce the gain in
the receiver and repeat the procedure.
5.
Repeat
through
for the D1 level using the D1 trim pot.
Note
Rosemount
™
has designed the circuits so that wherever saturation occurs (receiver or control
unit), the microprocessor always detects it. If the displayed detector levels cannot be set to
within this band or saturation cannot be avoided, optimize the detector levels ar the receiver.
Refer to
.
3.6
Source adjustments
Two trim potentiometers within the source unit allow adjustments to be made to the
intensity of the source and to the frequency of the chopper motor. Rosemount
™
sets these
at the factory, and they rarely need adjustment. Rosemount recommends that you consult
the company before making any adjustments within the source unit.
CAUTION!
DECREASED SOURCE LIFE
Increasing source intensity may severely reduce the source life.
3.6.1
Source intensity
Complete the following steps to adjust the source intensity in the analyzer.
1.
Loosen and slide the source unit cable gland so the cable can slip when the rear
cover plate is removed.
2.
Unscrew and remove the rear cover plate.
3.
) allows adjustment to the intensity of the source. Turn the trim
pot clockwise to increase the source intensity.
Configuration and startup
Reference Manual
33
Summary of Contents for Rosemount CCO 5500
Page 4: ......
Page 8: ...Contents iv Rosemount CCO 5500 ...
Page 28: ...System Wiring Diagram Figure 2 7 Install 20 Rosemount CCO 5500 ...
Page 32: ...Install 24 Rosemount CCO 5500 ...
Page 88: ...Operation 80 Rosemount CCO 5500 ...
Page 98: ...Maintenance 90 Rosemount CCO 5500 ...
Page 108: ...Troubleshooting 100 Rosemount CCO 5500 ...
Page 110: ...Spare parts 102 Rosemount CCO 5500 ...
Page 122: ...Return equipment to the factory 114 Rosemount CCO 5500 ...
Page 123: ...Return equipment to the factory Reference Manual 115 ...