12 V
Suppy rail for the plant status input.
+15 V, -15 V, and 0 VA
Supply rails for the analog current output.
VIN
Output from the D/A converter. 0 to 2.5 V repre-
sents 0 to 20 mA (4 mA at 0 V).
-15 V1, 0 V, and +15 V1
Isolated supply for the analog current output.
0 VB and +12 VB
Isolated supply for the plant status input.
F
Reference wave from the source unit (via the
power supply). Square wave, 30 to 45 Hz, 5 V
peak to peak, centered on 2.5 V.
Power supply
Description
0 V and +12 V
Power supply for the instrument.
Receiver
Description
0 V
0 V for the receiver.
T1
Detector output without conditioning.
T2
Detector output after first stage of gain.
T3
Detector output after both stages of gain.
Source
Description
S-
0 V supply to the heater cartridge
S+
12 V supply to the heater cartridge
M+ and M-
Supply to the chopper motor (+1 Vdc).
T3
Reference wave.
T4
Reference wave.
Troubleshooting
Reference Manual
99
Summary of Contents for Rosemount CCO 5500
Page 4: ......
Page 8: ...Contents iv Rosemount CCO 5500 ...
Page 28: ...System Wiring Diagram Figure 2 7 Install 20 Rosemount CCO 5500 ...
Page 32: ...Install 24 Rosemount CCO 5500 ...
Page 88: ...Operation 80 Rosemount CCO 5500 ...
Page 98: ...Maintenance 90 Rosemount CCO 5500 ...
Page 108: ...Troubleshooting 100 Rosemount CCO 5500 ...
Page 110: ...Spare parts 102 Rosemount CCO 5500 ...
Page 122: ...Return equipment to the factory 114 Rosemount CCO 5500 ...
Page 123: ...Return equipment to the factory Reference Manual 115 ...