Note
Rosemount
™
obtained factory settings under the following conditions:
•
2 m path length
•
Clean conditions
Calibration Data
Table 3-2:
Detector outputs
Factory
Config
D2
D1
E1
E2
Calibration data
Cal factor
Span factor
Output calibration
Set zero
Set span
3.8
Current output calibration
Now set up the current output using a calibrated multimeter set to DC current, 20 mA
max. Conduct this procedure as follows:
Note
Rosemount
™
sets the current output at the factory; do not alter it.
Procedure
1.
Connect the multimeter to the output terminals within the control unit terminals
+mA and -mA.
2.
Enter Mode 5
→
Configure Output
→
Fault Condition
→
Set Zero and adjust the level
using the arrow keys until
0 mA
is recorded. Record the value in brackets on the
display in
3.
Enter Mode 5
→
Configure Output
→
Fault Condition
→
Set Span and adjust the level
using the arrow keys until
20 mA
is recorded. Record the value in brackets on the
display in
Configuration and startup
50
Rosemount CCO 5500
Summary of Contents for Rosemount CCO 5500
Page 4: ......
Page 8: ...Contents iv Rosemount CCO 5500 ...
Page 28: ...System Wiring Diagram Figure 2 7 Install 20 Rosemount CCO 5500 ...
Page 32: ...Install 24 Rosemount CCO 5500 ...
Page 88: ...Operation 80 Rosemount CCO 5500 ...
Page 98: ...Maintenance 90 Rosemount CCO 5500 ...
Page 108: ...Troubleshooting 100 Rosemount CCO 5500 ...
Page 110: ...Spare parts 102 Rosemount CCO 5500 ...
Page 122: ...Return equipment to the factory 114 Rosemount CCO 5500 ...
Page 123: ...Return equipment to the factory Reference Manual 115 ...