A.3
Control unit
The control unit is a housed in a fully-sealed cast-aluminum enclosure. It houses the
microprocessor to monitor the data from the receiver and produces a 4-20 mA output
signal for gas levels within the flue.
A non-volatile RAM section, requiring no battery back-up, enables all of its operation data
to be retained during a power down condition. The instrument can resume operation
immediately when power is restored without having to be recalibrated.
Enter all operation data via a surface-mounted keypad. A 32-character LCD provides you
with measurement details and diagnostic information.
If required, inputs are available to receive the 4-20 mA outputs from the normalizing
measurement transducers (oxygen, temperature, and pressure). You can also enter this
data via the keypad or via the serial data port.
A serial communication facility within the processor allows the instrument to
communicate with other Rosemount
™
analyzers and a central data logging unit.
A.4
Power supply unit
The power supply unit is housed in a fully-sealed cast aluminum enclosure. A supply
voltage switch allows you to select from either 110 or 220 V. This provides an extremely
stable power source able to cope with large fluctuations in the supply voltage. Also, the
data valid and high gas alarm contact outputs are located in the unit.
A.5
Air purge
The air purge unit has an integral adjustable mount and provides the interface between
the site mounting flange and the source/receiver units. Rosemount
™
designed the air
purge to provide a steady laminar flow of air away from the instrument lens to prevent
optical contamination.
A supply of air to the purge is essential. See
for details.
A.6
Isolating valves
If you need to, you may attach isolating valves between the air purge units and the site
mounting flanges. Use the customer supplied isolating valves to provide protection for
personnel servicing instruments on high pressure ducts.
Theory of operation
Reference Manual
105
Summary of Contents for Rosemount CCO 5500
Page 4: ......
Page 8: ...Contents iv Rosemount CCO 5500 ...
Page 28: ...System Wiring Diagram Figure 2 7 Install 20 Rosemount CCO 5500 ...
Page 32: ...Install 24 Rosemount CCO 5500 ...
Page 88: ...Operation 80 Rosemount CCO 5500 ...
Page 98: ...Maintenance 90 Rosemount CCO 5500 ...
Page 108: ...Troubleshooting 100 Rosemount CCO 5500 ...
Page 110: ...Spare parts 102 Rosemount CCO 5500 ...
Page 122: ...Return equipment to the factory 114 Rosemount CCO 5500 ...
Page 123: ...Return equipment to the factory Reference Manual 115 ...