5.2
Corrective maintenance
The following maintenance procedures describe the necessary steps for removing and
replacing failed elements of the Rosemount
™
CCO 5500 Analyzer.
WARNING!
ELECTRIC SHOCK
Install all protective equipment covers and safety ground leads after equipment repair or
service. Failure to install covers and ground leads could result in serious injury or death.
5.2.1
Replace heater element
The heater element has a finite life and at some stage you will have to replace it.
Rosemount
™
designed the unit to give a minimum of two years continuous operation; you
can replace the heater element on site when necessary.
Replace a failed heater element according to the following procedure.
Procedure
1.
Shut down and lock out power to the Rosemount CCO 5500 Analyzer.
2.
Unbolt and remove the source unit from the air purge.
3.
Loosen the source unit cable gland (
) so the cable can slip when the rear
cover plate is removed.
Source Rear Cover Plate
Figure 5-2:
A. Rear cover plate
B. Cable gland
4.
Unsccrew and remove the rear cover plate.
5.
Unscrew and remove the circuit board (
).
Maintenance
82
Rosemount CCO 5500
Summary of Contents for Rosemount CCO 5500
Page 4: ......
Page 8: ...Contents iv Rosemount CCO 5500 ...
Page 28: ...System Wiring Diagram Figure 2 7 Install 20 Rosemount CCO 5500 ...
Page 32: ...Install 24 Rosemount CCO 5500 ...
Page 88: ...Operation 80 Rosemount CCO 5500 ...
Page 98: ...Maintenance 90 Rosemount CCO 5500 ...
Page 108: ...Troubleshooting 100 Rosemount CCO 5500 ...
Page 110: ...Spare parts 102 Rosemount CCO 5500 ...
Page 122: ...Return equipment to the factory 114 Rosemount CCO 5500 ...
Page 123: ...Return equipment to the factory Reference Manual 115 ...