If you use the
Keypad Input
, and the gas temperature is higher than 572 ° F (300 °C), the
compensation algorithm becomes less precise, and instrument accuracy deteriorates
accordingly. Rosemount does not recommend this.
Note
If normalization is not required, the instrument must hold the temperature of the gas in the duct
using the
Analog Input
option.
Press ENTER with the
Temperature
option selected to access
Set Standard Levels
and
Set Values
.
Use the arrow keys to toggle between these options and
Exit
.
Set Values
You can bring the normalizing data into the instrument in one of three ways:
•
Analog Input
This uses the 4-20 mA inputs within the processor to receive the measured
transducer data. The values at 4 mA and 20 mA will be requested should this option
be selected.
•
Serial Input
If you use an input unit, all normalizing data can be transmitted via the serial data
line.
•
You can enter a fixed value via the keypad. This is suitable where the value is stable
to about ±5%.
With an integrated system, set the lead analyzer's normalizing parameters to the 4-20 mA
inputs. Then set all the other analyzers to serial, and the normalization parameters are
transmitted down the serial data highway.
Oxygen
To correct the data to standard levels of oxygen, you must enter an estimate of the oxygen
at the point of measurement. If the oxygen level is being continuously measured, connect
the analog output of the oxygen analyzer into the Rosemount
™
CCO 5500 Analyzer and
select
Analog Input
. You must define this input as either
WET
or
DRY
depending on how the
Operation
70
Rosemount CCO 5500
Summary of Contents for Rosemount CCO 5500
Page 4: ......
Page 8: ...Contents iv Rosemount CCO 5500 ...
Page 28: ...System Wiring Diagram Figure 2 7 Install 20 Rosemount CCO 5500 ...
Page 32: ...Install 24 Rosemount CCO 5500 ...
Page 88: ...Operation 80 Rosemount CCO 5500 ...
Page 98: ...Maintenance 90 Rosemount CCO 5500 ...
Page 108: ...Troubleshooting 100 Rosemount CCO 5500 ...
Page 110: ...Spare parts 102 Rosemount CCO 5500 ...
Page 122: ...Return equipment to the factory 114 Rosemount CCO 5500 ...
Page 123: ...Return equipment to the factory Reference Manual 115 ...