Prerequisites
Note
Place the check cell at the receiver side.
Measurement conditions
For absolute verification, you must check the instrument when no measurement gas is
present. If a background concentration of measurement gas is present, an increase will still
be generated, but the net effect will be complex.
Mode
6
The ideal time to perform a check cell test is with the plant off, auto zero condition on, and
the analyzer well stabilized at zero. Do not insert the check cell in any other mode than
Mode
6
.
The analyzer does not update rolling averages while in Mode
6
.
Procedure
1.
Enter Mode
6
on the control unit before inserting the check cell into the analyzer.
Press ENTER when Mode
6
is displayed.
2.
Use an arrow key to toggle from
NO
to
YES
and press ENTER to access the check cell
function.
If you don't select this within five seconds, the instrument returns to the normal
OPERATING
mode.
3.
When you see this option, insert the check cell observing the procedure outlined
below.
4.
Remove the two screws retaining the cover on the check cell holder (
Operation
Reference Manual
77
Summary of Contents for Rosemount CCO 5500
Page 4: ......
Page 8: ...Contents iv Rosemount CCO 5500 ...
Page 28: ...System Wiring Diagram Figure 2 7 Install 20 Rosemount CCO 5500 ...
Page 32: ...Install 24 Rosemount CCO 5500 ...
Page 88: ...Operation 80 Rosemount CCO 5500 ...
Page 98: ...Maintenance 90 Rosemount CCO 5500 ...
Page 108: ...Troubleshooting 100 Rosemount CCO 5500 ...
Page 110: ...Spare parts 102 Rosemount CCO 5500 ...
Page 122: ...Return equipment to the factory 114 Rosemount CCO 5500 ...
Page 123: ...Return equipment to the factory Reference Manual 115 ...